How to read a peptide certificate of analysis

A well-made certificate of analysis is the researcher's best tool for knowing what is in the vial. This guide explains what each test measures (HPLC, mass spectrometry, content, water, endotoxins, solvents) and what it cannot measure.

Published October 9, 2026 · 7 min read · 15 sources

What a certificate of analysis is for

The certificate of analysis (COA) summarizes the tests performed on a specific batch. For a laboratory or a research group it is the first quality filter: peptide-related impurities can influence functional studies and lead to erroneous conclusions2. The EMA guideline, in effect since 1 June 2026, lists the expected specification, from identification and purity to counter-ion, residual TFA, water, residual solvents, bacterial endotoxins and microbiological purity5. That list is a good template for reading any certificate.

HPLC purity: what "≥98%" means

The most visible number usually comes from reversed-phase liquid chromatography with an ultraviolet detector, the routine quality-control method for peptides3. The calculation is area normalization: the area of the main peak against the area of all detected peaks. In a quality-control study from Ghent University, detection was at 210 and 215 nm, with a reporting threshold of 0.1% of the main-peak area1; detection at 220 nm has been described for oxytocin and acetates8. So "≥98%" means that at least 98% of the integrated ultraviolet signal belongs to the main peak.

That definition has limits worth knowing. Water and counter-ions are measured by separate tests and are not part of that percentage57. An impurity that elutes with the main peak is hidden inside it: the EMA asks manufacturers to consider the risk of co-eluting impurities and, when they appear as one peak, to apply the 1.0% qualification threshold5. A good certificate states the method and attaches the chromatogram with the integrated peaks; FDA rules require laboratory records to keep those data15.

What is in the remaining percentage? Mostly impurities typical of solid-phase synthesis: sequences missing an amino acid (deletions) or carrying an extra one (insertions), diastereomers from racemization, protecting-group adducts, oxidations and dimers2. Regulators set precise thresholds. Under the European Pharmacopoeia, peptide-related impurities are reported above 0.1%, identified above 0.5% and qualified above 1.0%5. The FDA, in its guidance on synthetic generics of glucagon, liraglutide, nesiritide, teriparatide and teduglutide (summarized in an official 2022 presentation), asks that any new impurity between 0.10% and 0.5% be identified and justified, and states that a new impurity "> 0.5% is not acceptable"6.

Identity: mass spectrometry

Purity tells how much of the material is a single component; it does not say which component. That is what mass spectrometry is for. Electrospray ionization (ESI) produces multiply charged ions: the spectrum of a large molecule appears as a series of peaks, each differing from its neighbor by one charge4. That is why a peptide of mass M shows up at mass-to-charge ratios such as (M + 2H)/2 or (M + 3H)/3, and the instrument software reconstructs the mass from that series. The EMA asks for tables of theoretical and observed mass values5. When reading a certificate, check that the observed mass matches the theoretical mass of the stated sequence, that it is clear whether monoisotopic or average mass is reported, and that the spectrum is attached.

Mass has a blind spot: two diastereomers weigh exactly the same, and racemization is a known synthesis impurity2. The EMA recommends at least two orthogonal identification methods (for example, mass together with relative retention time, peptide mapping, amino acid analysis or NMR) and notes that controlling diastereomers may require specific methods5. In the Ghent study, the main compound of one sample had a different structure from the one ordered1, an error that purity alone would not have revealed.

Purity is not content: counter-ions, salts and water

This is the point that causes the most confusion. A lyophilized powder with 99% HPLC purity is not 99% peptide by weight: it also contains counter-ions, water and sometimes solvents. The best-documented case is a candidate certified reference material of angiotensin II from the National Research Council Canada: the trifluoroacetate counter-ion accounted for nearly 25% of the mass and the final assigned content was 691 ± 9 mg/g7. In other words, about 69% of the powder was peptide. That value is the net peptide content; the EMA states that a chromatographic assay is expressed on the anhydrous, counter-ion-free substance5.

43 of 98peptides with a ≥95% certificate that met that purity in an independent check1
≈25%of the mass of an angiotensin II standard was TFA counter-ion7
0.10–0.5%range in which the FDA asks for new impurities to be identified and justified6

Why are there counter-ions? Peptides are cleaved from the resin and purified with trifluoroacetic acid, so cationic peptides are obtained as trifluoroacetate salts; most approved peptides, by contrast, are marketed as acetates8. The difference is not only about weight. In osteoblast cultures, TFA at 10⁻⁸–10⁻⁷ M reduced cell numbers, and the TFA salts of amylin and calcitonin masked a proliferative effect or suggested a nonexistent antiproliferative one; the authors recommended converting peptides to the hydrochloride before assessing their biological effects9. The counter-ion can be quantified by ion chromatography, and lyophilization with hydrochloric or acetic acid reduces TFA10. Salt forms also matter for GLP-1s, as explained in Compounded GLP-1s: salts, impurities and antibodies.

Water is the third component. Residual moisture is a critical quality attribute of lyophilized products, and Karl Fischer titration is the usual method for measuring it, with near-infrared spectroscopy as a nondestructive alternative11; the EMA includes it in the specification and expects it in stability studies when the powder is hygroscopic5. The practical consequence for the researcher is knowing whether the milligrams on the label refer to powder or to net peptide: for example, 10 mg of powder with 70% peptide content provide 7 mg of peptide. How to store that powder afterwards is covered in Peptide stability.

Endotoxins, sterility and residual solvents

The endotoxin test and the sterility test measure different things. Endotoxins are determined with the Limulus amebocyte lysate (LAL) assay, which the FDA has accepted for more than 30 years in place of the rabbit pyrogen test and which is described in USP chapter <85>12. The result should be reported with a number and units, for example in EU/mg, alongside its limit. A Hungarian study of semaglutide sold online shows why both tests matter: the three lyophilized vials purchased had no viable microorganisms at the time of testing, yet all contained endotoxins, between 2.16 and 8.95 EU/mg13.

Residual solvents are governed by the ICH Q3C guideline, which places them in three classes: solvents to be avoided (class 1), solvents to be limited (class 2) and solvents with low toxic potential (class 3)14. For class 2 it sets specific limits: 410 ppm for acetonitrile, 600 ppm for dichloromethane and 880 ppm for N,N-dimethylformamide; for class 3, up to 0.5% is considered acceptable without further justification14.

Batch, dates and laboratory

A certificate is valid only for the batch it describes. FDA good manufacturing practice rules require laboratory records to identify the sample with its lot number, the date it was taken and received, the method used, all data obtained (including instrument graphs, charts and spectra identified with the lot), the calculations, the initials or signature of whoever performed each test with the date, and review by a second person15. The EMA expects material traceability from synthesis through to the final active substance5.

In-house or third-party testing? The Ghent group found that although all 98 peptides came with a certificate stating ≥95% purity, only 43 met it, and recommended that researchers maintain their own quality control1. For a laboratory, the strongest option is to combine the supplier's certificate with independent verification, at least of identity by mass, for the key batches of each research protocol.

Relying on the certificates of analysis as provided by the supplier might have serious consequences for peptide research
Verbeke et al., Ghent University1

Red flags and the limits of a certificate

  • No mass spectrum or observed mass: purity alone does not confirm identity15.
  • Identical chromatograms (same retention times, same areas, same baseline) across different batches: each batch generates its own data15.
  • No lot number, no test dates or no signatures15.
  • Purity without peptide content, counter-ion or water when the material is weighed to prepare solutions57.
  • Endotoxins reported as "conforms" with no number or units1213.
  • A 99.9% figure with no integrated chromatogram or method, or a laboratory that cannot be identified115.

Even a flawless certificate has limits. It is a snapshot of the batch on the test date; that is why the EMA requires separate stability studies5. Market studies also show how far the declared can be from the measured: in three vials of semaglutide labeled at 99%, measured purity was 7.7–14.4%13, a finding reviewed in What is really in the vial?. A well-read certificate does not remove those risks; it lets you detect them in time.

Sources

  1. Verbeke F, Wynendaele E, Braet S, D'Hondt M, De Spiegeleer B. Quality evaluation of synthetic quorum sensing peptides used in R&D. J Pharm Anal. 2015;5(3):169-181. PubMed 29403929 ↗
  2. D'Hondt M, et al. Related impurities in peptide medicines. J Pharm Biomed Anal. 2014;101:2-30. PubMed 25044089 ↗
  3. Stalmans S, et al. Quality control of cationic cell-penetrating peptides. J Pharm Biomed Anal. 2016;117:289-297. PubMed 26397208 ↗
  4. Fenn JB, Mann M, Meng CK, Wong SF, Whitehouse CM. Electrospray ionization for mass spectrometry of large biomolecules. Science. 1989;246(4926):64-71. PubMed 2675315 ↗
  5. European Medicines Agency. Guideline on the development and manufacture of synthetic peptides (EMA/CHMP/CVMP/QWP/367182/2025). 4 December 2025; in effect 1 June 2026. ema.europa.eu ↗
  6. Pang E. U.S. Food and Drug Administration, SBIA 2022. Assessing immunogenicity risk of peptides: the synthetic peptide guidance (ANDAs for certain highly purified synthetic peptide drug products that refer to listed drugs of rDNA origin) and PSGs. fda.gov ↗
  7. Melanson JE, et al. Purity assignment for peptide certified reference materials by combining qNMR and LC-MS/MS amino acid analysis results: application to angiotensin II. Anal Bioanal Chem. 2018;410(26):6719-6731. PubMed 30143839 ↗
  8. Sikora K, Jaśkiewicz M, Neubauer D, Migoń D, Kamysz W. The role of counter-ions in peptides: an overview. Pharmaceuticals (Basel). 2020;13(12):442. PubMed 33287352 ↗
  9. Cornish J, et al. Trifluoroacetate, a contaminant in purified proteins, inhibits proliferation of osteoblasts and chondrocytes. Am J Physiol. 1999;277(5):E779-E783. PubMed 10567002 ↗
  10. Mrozik W, et al. Determination of counter-ions in synthetic peptides by ion chromatography, capillary isotachophoresis and capillary electrophoresis. J Pept Sci. 2012;18(3):192-198. PubMed 22252914 ↗
  11. Clavaud M, et al. Near-infrared spectroscopy to determine residual moisture in freeze-dried products: model generation by statistical design of experiments. J Pharm Sci. 2020;109(1):719-729. PubMed 31499067 ↗
  12. U.S. Food and Drug Administration. Pyrogen and endotoxins testing: questions and answers (Edition 2). Guidance for industry. fda.gov ↗
  13. Ashraf AR, et al. Multifactor quality and safety analysis of semaglutide products sold by online sellers without a prescription: market surveillance, content analysis, and product purchase evaluation study. J Med Internet Res. 2024;26:e65440. PubMed 39509151 ↗
  14. International Council for Harmonisation / European Medicines Agency. ICH Q3C (R9) guideline on impurities: guideline for residual solvents (EMA/CHMP/ICH/82260/2006). 2024. ema.europa.eu ↗
  15. U.S. Code of Federal Regulations. 21 CFR 211.194, Laboratory records. ecfr.gov ↗

Informational article. It summarizes news, official statements and published studies; it is not medical or legal advice. Sources were checked on the publication date.